Status: UPDATED | Advisory ID: CVE-2019-25718
| CVE | CVE-2019-25718 |
| CVSS Score / Version | 8.4 (High) / CVSS v3.1 |
| Updated | 2026-07-22 |
| CVSS Vector | CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H |
| CVSS Prose | attack vector is local; attack complexity is low; privileges required is none; user interaction is none; scope is unchanged; confidentiality impact is high; integrity impact is high; availability impact is high. |
| Affected products | draeger infinity_explorer_c700_firmware |
| Classified as | CWE-451 (User Interface (UI) Misrepresentation of Critical Information) |
| Vendor | Product | Affected Versions | Patch Status |
|---|---|---|---|
| draeger | infinity_explorer_c700_firmware |
| Subsystems | General OT |
| Sectors | Multiple |
Dräger Infinity Explorer C700 contains a privilege escalation vulnerability that allows attackers to break out of kiosk mode and access the underlying operating system through a specific dialog interaction. Attackers can exploit this kiosk escape to take control of the operating system and cause the device to display incorrect or no information from the connected Delta Family patient monitor. (NVD)
Monitor draeger's web page for any future patch releases. See vendor advisory link below.