Status: NEW | Advisory ID: CVE-2019-25719
| CVE | CVE-2019-25719 |
| CVSS Score / Version | 8.6 (High) / CVSS v3.1 |
| Updated | 2026-07-22 |
| CVSS Vector | CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H |
| CVSS Prose | attack vector is network; attack complexity is low; privileges required is none; user interaction is none; scope is unchanged; confidentiality impact is low; integrity impact is low; availability impact is high. |
| Classified as | CWE-924 (Improper Enforcement of Message Integrity During Transmission in a Communication Channel) |
Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions. Attackers with access to an enabled Infinity network port or physical proximity to a wireless access point can modify device settings such as alarm states or alarm limits, and overwhelm the system with incoming data causing the device to reboot and lose network functionality. (NVD)
| Source | Reference |
|---|---|
| NVD | https://nvd.nist.gov/vuln/detail/CVE-2019-25719 |
| CVE | https://www.cve.org/CVERecord?id=CVE-2019-25719 |