Status: NEW | Advisory ID: CVE-2019-25722
| CVE | CVE-2019-25722 |
| CVSS Score / Version | 7.6 (High) / CVSS v3.1 |
| Updated | 2026-07-22 |
| CVSS Vector | CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H |
| CVSS Prose | attack vector is adjacent; attack complexity is low; privileges required is none; user interaction is none; scope is unchanged; confidentiality impact is low; integrity impact is low; availability impact is high. |
| Classified as | CWE-798 (Use of Hard-coded Credentials) |
Dräger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contain hard-coded plaintext credentials in source code and a denial-of-service vulnerability that allows local and remote attackers to compromise device integrity across all software versions. A local attacker with direct device access can use the hard-coded credentials to access service and clinical accounts and alter device configuration, while a remote attacker can send malformed network packets to cause repeated device reboots, ultimately resulting in loss of network connectivity and disruption of patient monitoring. (NVD)