Dräger Infinity Acute Care System and M540 Patient Monitors Network Message Handling Flaw Risks Data Integrity Across Multiple Software Versions

Last refreshed2026-11-17

Status: UPDATED  |  Advisory ID: CVE-2022-4992

Key Details

CVECVE-2022-4992
CVSS Score / Version
8.6 (High) / CVSS v3.1
Updated2026-07-22
CVSS Vector
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H
CVSS Prose
attack vector is network; attack complexity is low; privileges required is none; user interaction is none; scope is unchanged; confidentiality impact is low; integrity impact is low; availability impact is high.
Affected products
Dräger Infinity Acute Care System and Dräger Standalone Infinity M540 patient monitor
Classified as
CWE-345 (Insufficient Verification of Data Authenticity)
Exploitation prediction (EPSS)
0.16% probability of exploitation in the next 30 days (5% percentile) -- FIRST.org's EPSS model.

Affected Products, Subsystems & Sectors

VendorProductAffected VersionsPatch Status
DrägerInfinity Acute Care System
DrägerStandalone Infinity M540 patient monitor
SubsystemsGeneral OT
SectorsMultiple

What to Know

Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors versions VG4.1.1, VG4.0.3, and lower (with VG4.2 partially affected) contain a network message handling vulnerability that allows remote attackers to inject spoofed or tampered data and cause denial-of-service conditions. Attackers can compromise network communications to modify device settings such as alarm states or alarm limits, or overwhelm the system with excessive network traffic causing the Cockpit or M540 to reboot and lose network functionality. (NVD)

What to Do

Monitor Dräger's web page for any future patch releases.

https://static.draeger.com/security/download/PSA-19-255-03_Product-Security-Advisory-IACS.pdf

References

SourceReference
NVDhttps://nvd.nist.gov/vuln/detail/CVE-2022-4992
CVEhttps://www.cve.org/CVERecord?id=CVE-2022-4992
Vendor advisoryhttps://static.draeger.com/security/download/PSA-19-255-03_Product-Security-Advisory-IACS.pdf
Report compilation © 2026 OT-Advisories.com  ·  CC BY 4.0  ·  Source materials remain the property of their respective publishers as cited.